Quantify small molecules, peptides, biosimilars, and biologics using the Thermo Scientific TSQ Vantage triple quadrupole mass spectrometer. With high sensitivity and low noise, the TSQ Vantage system delivers high assay precision and accuracy for a range of bioanalytical, clinical research, environmental, and proteomics applications.

 

Manufacture Specification 

Mass Range

  • m/z 10-1500 daltons (Da)

Resolution

  • 7500 (FWHM) at m/z 508 of polytyrosine
  •  Resolution is continuously adjustable to better than 0.1 Da peak width (FWHM) across the entire mass rang

Mass Stability

  • Mass assignment will be within ±0.050 Da over a 24-hour period. The laboratory room temperature must be maintained between 15-27°C (59-81°F). The optimum temperature of operation is between 18-21°C (65-70°F). The room temperature may not change by more than 5°C (9°F) during this period

Gas

  • Collision gas: 99.995% pure Argon 
  • Collision gas supply pressure: 135 ±70 kPa (20 ±10 psig) 
  • Sheath/aux/sweep gas: 99% pure Nitrogen 
  • Sheath/aux/sweep gas supply pressure: 690 ±140 kPa (100 ±20 psig) 
  • Maximum sheath gas consumption: ~ 20 L/min

Additional information

Weight 395 lbs
Dimensions 48 × 40 × 30 in
Make Thermo Scientific
Model TSQ Vantage
Serial TQU02828
Working Working
Item Condition Used
Includes Power Cord
Warranty 30-Day Warranty, 100% Parts and Labor
Voltage 220V
hertz 50Hz/60Hz